EXTRACTIVE GAS ANALYZER

BeamCell™

Real-time extractive gas analysis delivering high-precision data rapidly, allowing multi-point sampling and analysis.

The BM-H-3 BeamCell is a state-of-the-art turnkey solution utilizing Tunable Diode Laser Absorption Spectroscopy
(TDLAS) for precise analysis of extracted/sampled gas. Equipped with push-in gas connectors (G1/8) adapted for
6-mm or 8-mm gas tubing, it ensures seamless integration.

Effortless integration: By extracting gas in a parallel flow for analysis within the controlled environment of the
BM-H-3 BeamCell, very high measurement accuracy is achieved. The analyzer is easily integrated into industrial
environments, with standard communication interfaces compatible with most PLC systems.

Unique compact form factor: The compact design of the BM-H-3 BeamCell makes it suitable for both new
installations and retrofits into existing infrastructure, offering flexibility and ease of integration.

High-speed sampling: The fast response time of the BM-H-3 BeamCell provides true real-time analysis data. The
speed of the analyzer also allows for sampling of multiple points in sequence within a few seconds.

Advanced TDLAS technology: By leveraging the robust and proven TDLAS technology, the BM-H-3 BeamCell
delivers rapid (100 µs) and highly sensitive (ppb-level) gas analysis. The technology is inherently self-referencing and
calibration-free, ensuring minimal maintenance.

Versatile gas analysis: The analyzer is capable of detecting a wide range of gases commonly found in industrial and
laboratory settings, including greenhouse gases such as methane (CH₄). Other commonly analyzed gases include
O₂, CO₂, CO, NH₃, HF, H₂S, H₂S, H₂O and HCl.

Energy efficient design: The highly energy-efficient design allows the BM-H-3 BeamCell to be integrated in portable
systems capable of running on battery power for hours in mobile analysis applications.

Durable design: The analyzer is designed for heavy industrial use and is constructed to withstand harsh and
corrosive gases. The flow chamber in the BM-H-3 BeamCell is capable of withstanding harsh chemicals such as
sulfuric acid, ensuring long-term reliability ain demanding applications.

• Turnkey extractive gas analyzer
• Absolute gas quantification
• Calibration-free gas sensing
• Acid resistant materials in flow-chamber
• Digital and analog data interfaces
• IP67 grade mechanics
• Easy service and maintenance

Examples of applications:

  • Extractive Gas Analysis

  • Laboratory Gas Sensing
  • Industrial Process Monitoring
  • Explosion Risk Management
  • Mobile Analysis in Harsh Environments
GasAnalyzer precision (ppm)*)
O₂30
HF0.05
CO1
CO₂2.5
CH1
H₂S1.5
NH₃1
H₂O1

*) Under standard test conditions: L = 0.2 m, t = 1 s, P = 1 atm, T = 300 K, largest of 1% relative and specified precision.

ParameterSymbolMinTypicalMax
Ramp periodTR100 µs120 µs240 µs
Temperature set stability0.1 mK
Laser currentIR0 mA400 mA
Data sampling channels1
Data sampling resolution24 bits
Data sampling noise (V)0.1 µVrms*
Data sampling noise (ADU)0.1 *
Low-light limit (10 s)1 nW
Gas connector threadingG1/8
Gas tube dimension6 mm / 8 mm
*1 s sampling time

 

Interfaces
InterfaceModelMountedQuantity
USB53398-0471 Communication, DataYes2
USBMini USB, Firmware upgradeYes2
RS-4854 wire Full Duplex – protectedYes1
RS-4854 wire Full Duplex – service dataYes1
RS-485Half DuplexYes1
Trig In4-30 VYes2
IO0-12 VYes2
IO Supply12 VYes1
Relay OutputG6K-2F-5DC, NC/NOYes2
4-20 mAPassive / ActiveExpansion2
Expansion connector2C, SPI, GPIO, ADC, Loop Relay, Sync, GND, PWM, UART, Relay, IO, 5 V, 2.5 V, 12 V, PWR_IN, 4-20 mA, 0-10 VYes3
Sync signalsDaisy chain configurationYes2
Master clk in / out73412-0110Yes2

 

Electrical Characteristics
ParameterSymbolMinTypicalMax
Supply voltageVin15 VDC*24 VDC32 VDC
Power consumption5 W**
TEC driver power0 W3.56 W
Comm. link length30 m
Startup-time (ambient)tsu5 s
*Degraded noise performance if Vin < 15 V | **50 mA laser diode

 

Other
ParameterSymbolMinTypicalMax
Operating temperatureTop-10 °C50 °C
Humidity (non-condensing)40% @ 50 °C / 80% @ 30 °C
IP classificationIP67
Infrared laserLaser Class I
CE-marked EU directives2014/35/EU, 2012/19/EU, 2011/65/EU, EN61000-6-2:2005, EN61000-6-2:2019, EN61000-6-4:2007, EN61000-6-4:2019

Download BeamCell™ Datasheet

Product Images

BOOK A FREE DEMO

Interested in learning more?

Book a free demo or contact us to learn more about our products and technology as well as explore how your gas analysis applications can be solved with our next-generation of TDLAS gas analyzers.